Customization: | Available |
---|---|
Luna 16V: | Veterinary CT Scanner |
Transport Package: | by Air or by Sea |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make at Made-in-china.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Aperture | 70cm |
One-button positioning | Preset 3 protocols |
Scan speed/360° | 0.7*,0.8*,0.9,1.0,1.5,2.0s |
Scan FOV | 43cm |
Tilt range | Digital tilt±45° (step0.1) |
Auto voice | Support |
Max.horizontal travel range | 1600mm |
Horizontal scannable range | 50mm~1500mm |
Horizontaltravel speed | 1~200mm/s |
Vertical table travel range | 522mm~878mm |
Max.table load | 206kg |
One-key patient table release | Support |
Patient table cradle switches | Support |
Anode heat capacity | 3. 5MHU |
Equivalent anode heat capacity | 9.5MHU(with iDream) |
Cooling rate | 735kHU/min |
Focal spot size | Large: 1.2mm×1.4mm Small: 0.7mm×0.&mm |
Power rating | 42kW |
Equivalent power rating | 113kW (with iDream) |
kV settings | 70,80,100,120,140kV |
mA range (Step size) | 10~350mA (1mAstep) |
Material | Solid-state GOS |
No. of Detector rows | 24 rows |
Max. number of slices/rotation | 32 (conjugated mode) |
No.of detector channels per row | 768 |
Total No.of detector elements | 18432 |
Min.slice thickness | 0.6mm |
Detector width | 19.2mm |
Max.data sampling rate | 4800 views/360° |
Scout scan | Supports 3 modes: AP.lateral and dual; Scannable range 50~1500mm; |
Acquisition modes | 32×0.6mm 32×1.2mm 16×0.6mm 16×0.1.2mm |
Min slice thickness | 0.6mm |
Collimation width selection | 19.2mm,9.6mm |
Pitch factor | 0.3~1.75 (Multiple selections) |
Max.continuous scan time | 100s |
Recon FOv | 50~430mm; 50~600mm (Extended). |
Recon matrices | 512×512,768x 768,1024×1024 |
Recon speed | ≥20ips |
Display matrix | 1024×1024 |
Metal artifact reduction | Standard |
Beam hardening artifact reduction | Standard |
Partial volume artifact reduction | Standard |
Steaking artifact reduction | Standard |
Helical scan artifact reduction | Standard |
Motion artifact reduction | Standard |
Spatial resolution | ≥16 lp/cm @ 0% MTF;X-Y plane ≥15 lp/cm@0% MTF;Z plane |
Low-contrast resolution | [email protected]%@25mGy; |
lmage Noise | ≤0.35%(Central dose≤40 mGy) |
CT HU scale | Standard: -1024HU ~+3072HU Extended: -32768HU~+32767HU |
CPU | lntel Xeon 4 core,4 threads, frequency 2.9GHz,cache 8.25MB |
RAM | 16GB ECC |
Hard disk | 3TB(system disk 0.3TB+image disk 0.7TB + raw data disk 2TB) |
Monitor | Size: 24 inch, LCD Resolution: 1920×1200 Brightness: 600cd/m2 Contrast 1000:1 |
lmages storage | ≥1,300,000 images(512×512) |
External storage | DVD/CD RW, USB |
Printing interface | DICOM 3.0 standard |
Dedicated pediatric protocols | Standard |
Auto-mA | Standard |
V-Dose check | Standard |
Low dose lung screening | Standard |
240° exposure | Standard |
V-Beam | Standard |
V-Dose report | Standard |
iDream lterative reconstruction | Standard |
V-Bolustracking | Standard |
V-Bolus timing | Standard |
System | Length | Width | Height | Weight |
Gantry | 1970mm | 940 mm | 1850 mm | 1200 kg |
Table | 580 mm | 2400 mm | 950 mm | 350 kg |
Console | 400 mm | 630 mm | 630 mm | 50 kg |
Distributor | 810 mm | 435 mm | 665 mm | 200 kg |
Scanning room dimension | Min.area: 18m2 (5000mm x 3600mm) Recommended room size: 30m2 (6000mm x5000mm) |
Temperature&Humidity | Temperature: scanning room: 20~26°C;operating room: 18~28ºC Humidity: scanning room: 30%~70%,no condensation; operating room: 20%~80%,no condensation |
Power supply requirements | Power capacity: 70kVA Power supply option: 3 phase 380 VAC, voltage variation: tolerance ≤+10% Frequency: 50 Hz or 60 Hz, tolerance ≤±1Hz |
Intelligent energy saving | Insitum series CT scanners are designed to be energy saving,and have further optimized standby mode, which reduces the live operation of high-voltage control and data acquisition devices.They only keep necessary components in working state. This does not affect normal start-up efficiency, yet annual power consumption is reduced by 2815kW-h when the device is turned on for 10 hours a day, 6 days a week,which is 62.5% lower than the earlier design. |